Maintenance; Replacing The Current Input Fuse; Replacing The Power Supply Fuse - Presys PCON-Y17 Series Technical Manual

Pressure controller
Table of Contents

Advertisement

PRESYS
 Instruments
MEASure[:SCALar]:VOLTage[:DC]?
Description: Voltage Input (Auxiliary) Reading.
Parameters: ---
Response: V Reading
OUTPut:MODE <mode>
Description: Sets the Pressure Operation Mode.
Parameters: VENT | CONTrol | MEASure
Response: ---

3 - Maintenance

3.1. Replacing the Current Input Fuse

To replace the PCON-Y17 mA Input (Auxiliary) fuse, proceed as follows:
- Rotate the fuse holder counterclockwise (see see Fig. 01 for DT Version or Fig. 03 for
Rack Mounting Version – Fuse mA input). Try to use a plastic tool to avoid damaging the
fuse holder;
- The fuse holder will be released;
- Pull the fuse holder and remove the fuse;
- Place the spare fuse (Time-Lag Type Fuse 32 mA / 250 V / 5 x 20mm).

3.2. Replacing the Power Supply Fuse

In case of blow, the 1 Amp fuse (Time Delay Type Fuse 1 A / 250 V / 5x20mm),
placed within the power supply socket on the rear can be replaced.
The fuse may blow due to a voltage spike in the mains or a calibrator component
fault. Replace the fuse once. If a second fuse blows again, it is because the fault is not
that simple. In this case, contact the Presys technical support.
Page 60
PCON-Y17

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pcon-y17-dtPcon-y17-rmPcon-y17

Table of Contents