f) Step Function
The STEP programming makes the calibrator output vary in pre-defined steps. It
is useful in calibrations where some scale points are verified; for example 0% - 25% -
50% - 75% - 100%.
To generate setpoints, you must set the minimum and maximum range points
(MIN and MAX) and the desired number of points.
If you want each step is achieved automatically after a preset time, you must set
the TIME (in seconds). In this case, the steps are controlled automatically and
continuously.
To start the STEP function, change the status of the EXEC switch to ON. Once
the function is activated, the output starts executing the STEP program. The steps are
automatically and continuously controlled according to the set time.
Fig. 24 - STEP Function (automatic step program)
If a zero value is set for the time, the steps are not changed automatically (manual
step program). In this case, to move to the next step you must press the next set point
value button. By pressing the previous set point button, it returns a step.
Page 22
Fig. 25 - STEP Function (manual step program)