RM Version – Rack Mounting
Note:
For the 3000 psi Pressure Control range there is no VENT PORT.
On the PCON-Y17-RM, the VENT and SUPPLY ports of the pressure control module
are located on the back of the module, as well as the power supply (100 to 240 Vac) and
Ethernet port.
2 - Calibrator Operation
Turn on the PCON-Y17 by pressing the power switch (located on the rear panel
for the DT version, and on the front panel for the RM version).
When powered on, the calibrator goes through a self-test routine. In case of
failure, it displays a message to indicate the error; if that occurs, please contact the
Presys Technical Assistance department.
After the self-test is completed, the display shows the main menu, as showed
below.
WARNING!
During startup, do not connect any instrument to the output port of the
Pressure Control Module. Before making connections to the Pressure
Control Module, enter the CALIBRATOR menu and wait for the automatic
VENT operation.
The main menu is divided in the following functions:
Fig. 03 – Front Panel PCON-Y17-RM
Fig. 04 - Main Menu
PCON-Y17
Page 7