The result will be displayed on the screen, and a progress bar is displayed to
indicate the calibration remaining time. At the end of the calibration, a report is shown
with the generated values, the obtained values, the expected values, and the error. If the
error is higher than the registered value for the task, the line appears in red.
The first time that a task is performed, it will be saved as As-found (before
adjustment). If it runs again, it will be saved as As-left (after adjustment). The results are
saved in the calibrator and can be viewed at any time.

2.3.3. Viewing Results

Once a task has been performed, it remains saved in the calibrator.
To view the results of a calibration by the PCON-Y17, select TASKS in the main
menu.
Enable the ● PERFORMED option. The list will show only the tasks that have
been performed. Select the desired work order and press OK. On the screen, the report
with the calibration points, the obtained values, expected values and the errors will be
shown. If the error is higher than the value registered for the task, the line appears in red.
Fig. 71 - Viewing the Task Result in the Calibrator
The icon
Calibrator. To access the files saved on the PCON-Y17 internal memory, connect the
USB cable to the computer (type A USB) and to the PCON-Y17 (Micro-B USB, see
Fig. 01 for DT Version or Fig. 03 for Rack Mounting Version).
To save the data in a Pen Drive or External HD, press the Pen Drive icon
saving the data.
To print the Calibration Report, press the printer icon
been previously configured in SETTINGS > SYSTEM > PRINTER CONFIG and must be
Page 52
saves the task data in a PDF file in the internal memory card of the
PCON-Y17
after
. The printer must have