PRESYS
Instruments
When you want to load the saved configuration into an instrument, go to the
DEVICE> DOWNLOAD / VIEW menu. To select the configuration file, double click on it.
Press the WRITE button to download the configuration file for the connected
instrument. Before the instrument is fully configured, some confirmation messages will
be displayed. To cancel, press X. To continue, press OK. At the end of the configuration,
the Configuration Write Complete message appears.
2.3. Automatic Tasks
In PCON-Y17, automatic calibration can be created and performed tasks. This
option can be used to generate calibration work orders for transmitters and indicators.
PCON-Y17 also creates and performs tasks for calibration of analog gauges, using the
inverted calibration method, so the user can change the control set points in small steps
until the expected value is being indicated on the gauge.
2.3.1. Creating Tasks
To create tasks from the main menu, select CALIBRATOR. Select the desired
type of Input (Auxiliary) and the Pressure Control Module configuration (select the
Control mode). For example, to calibrate a pressure transmitter, select the Control mode
and mA auxiliary input (which will be connected to the current output of the transmitter).
For a pressure indicator, e.g., selected NONE for the auxiliary input.
To calibrate an analog gauge using the inverted calibration method, select NONE
for the input.
Press the
Page 48
Fig. 65 - Saving the Configuration of a Device
icon, and select TASKS and CREATE NEW TASK.
PCON-Y17