80
FEP630, FEH630 ELECTROMAGNETIC FLOWMETER | OI/FEP630/FEH630-EN REV. D
... 8 Operation
... Parameter descriptions

Menu: Device Info

This menu is only used to display the device parameters. The parameters are displayed independently of the configured access level,
but cannot be changed.
Menu / parameter
Device Info
...Sensor
...Transmitter
Device Info / ...Sensor
QmaxDN Type
Meter Size
Qm Max DN
Qv Max DN
Mains Frequency
Exitation Freq
Coil Current Regul.
Set Point Curr.
DC feedback reg.
Sensor ID
Sensor Serial No.
Sensor Run Hours
...Se Calibration
Sensor TFE Function
Device Info / ...Sensor / ...Se Calibration
Se Span
Se Zero
First Cal. Date
Last Cal. Date
Cal. Cert. No.
First Cal. Location
Last Cal. Location
Description
Selection of submenu '...Sensor' using
Selection of submenu '...Transmitter' using
For informational purposes only.
Nominal diameter of sensor.
The value is the maximum mass flow at a flow velocity of 10 m/s.
The value is automatically set through the selected nominal diameter, multiplied by the set density.
The value provides the maximum volume flow at a flow velocity of 10 m/s.
The value is set automatically via the selected nominal diameter.
Supply frequency for the power supply.
Frequency used to operate the magnetic coils of the flowmeter sensor.
For service information only.
Current used to operate the magnetic coils of the flowmeter sensor.
For service information only.
ID number of the sensor.
Serial number of the sensor.
Operating hours of the sensor.
Selection of submenu '...Se Calibration' using
Shows if the total filling electrode (TFE) has been activated or deactivated.
Calibration value in the forward flow (direction) and reverse flow (direction) of the sensor.
Date of first calibration of sensor (calibration of new device).
Date of last calibration of sensor.
Identification (number) of the relevant calibration certificate.
Place of first calibration of the sensor.
Place of last calibration of sensor.
.
.
.