Error Messages On The Lcd Display; Switching To The Configuration Level (Parameterization) - ABB ProcessMaster FEP630 Series Operating Instruction

Electromagnetic flowmeter
Hide thumbs Also See for ProcessMaster FEP630 Series:
Table of Contents

Advertisement

64
FEP630, FEH630 ELECTROMAGNETIC FLOWMETER | OI/FEP630/FEH630-EN REV. D
... 8 Operation
... Switching to the information level

Error messages on the LCD display

In the event of an error, a message consisting of a symbol and
text (e.g. Electronics) appears at the bottom of the process
screen.
The text displayed provides information about the area in which
the error has occurred.
Process display
Electronics
The error messages are divided into four groups in accordance
with the NAMUR classification scheme. The group assignment
can only be changed using a DTM or EDD:
Symbol
Description
Error / failure
Function check
Outside of the specification
Maintenance required
The error messages are also divided into the following areas:
Range
Description
Operation
Error / alarm due to the current operating
conditions.
Sensor
Error / alarm of the flowmeter sensor.
Electronics
Error / alarm of the electronics.
Configuration
Error / alarm due to device configuration.
Note
For a detailed description of errors and troubleshooting
instructions, please see Diagnosis / error messages on
page 109.
Switching to the configuration level
(parameterization)
The device parameters can be displayed and changed on the
configuration level.
Process display
1. Switch to the configuration level with
Access Level
Read Only
Standard
Service
Back
Select
2. Select the desired level of access using
3. Confirm the selection with
Note
There are three levels of access. A password can be defined for
level 'Standard'.
There is no factory default password. For security reasons it
is recommended to set a password.
The password prevents access to the parameterization via
the buttons on the device. For further access protection via
DTM or EDD (HART®, PROFIBUS®, Modbus®) the hardware
write protection switch must be set (see Hardware settings
on page 52).
Access Level
Description
Read Only
All parameters are locked. Parameters are read only and
cannot be modified.
Standard
All the parameters can be changed.
Service
Only ABB Customer Service has access to the Service
menu.
.
/
.
.