Important information in advance Congratulations! By purchasing this Mass Flow Meter FMT200-ECO2 (Sensyflow eco2), you have acquired a device which satisfies the highest quality and performance standards. FMT200-ECO2 (Sensyflow eco2) is the result of our long experience; it has been conceived in direct response to customer requirements.The modular set- up and the many adaptation possibilities guarantee utmost ease of operation and great versatility.Our cal-...
The examination intervals must be chosen in such a way that any damages that can be expected can be recognised in time. The examinations must be performed at least every three years. The examinations can be skipped if the electrical system is continuously monitored by a responsible en- gineer. Thermal Mass Flowmeter FMT200-ECO2 42/14-41 EN...
EC declaration of conformity (ATEX) . For the operation of Zone2 / Zone 22 certified devices only the values written down in this declaration of conformity (ATEX) are relevant. Warning Operating the device in Zone 1/ 21 or Zone 0 / 20 is not permitted. STOP 42/14-41 EN Thermal Mass Flowmeter FMT200-ECO2...
Description 3 Description 3.1 Process connectors The process connectors provided for FMT200-ECO2 (Sensyflow eco2) ensure user-specific utilization without extra mounting costs. Furthermore, they enable easy installation and removal of the measurement sensor for maintenance and cleaning purposes. 3.2 Measuring sensor Apart from the sensor unit, the measuring sensor, which is set up as a pipe, also contains the transformer electronics.
• Components affecting flow, such as valves or shut-off fittings should be possibly installed behind the measuring point. Should there be any uncertainty or questions regarding the measuring point to be selected, please con- tact the ABB Service Team in Alzenau. Tel: +49 6023/92-3267 or +49 6023/92-3597...
Connection to the small flange adapter 4.4.2 G1“, G3/4“ threaded joints 4.4.3 G1/2“, G3/8“ threaded joints Both adapters are equipped with a High-Tec flow conditioner sieve on the inflow side. Because of it, no steadying lengths are required. Thermal Mass Flowmeter FMT200-ECO2 42/14-41 EN...
Installation and start-up 4.4.4 Legris hose adapters The hose adapters are screwed into the G3/8" process adapters. 4.4.5 Transair adapters The Transair adapters are screwed into the G 3/4" adapters. 42/14-41 EN Thermal Mass Flowmeter FMT200-ECO2...
Upon selection, the analog output of the current output supplies an active signal of 0(4)...20 mA, i.e. the FMT200-ECO2 (Sensyflow eco2) device supplies the current independently. For this reason, do not use a 2-wire power supply unit or an active input of a PLC, but rather a passive signal receiver.
Sensyflow eco1 and FMT200-ECO2 (Sensyflow eco2) are compatible. Using the appropriate electrical adapter, FMT200-ECO2 can be connected to existing plants. As "interface" and "digital output" function- ality is not available with Sensyflow eco1, there is no wiring within the adapter for these functions.
Fig. 6-2 Verzeichnis = Directory Click on the button „Verzeichnis wechseln“ to change the installation directory Setup beenden (Exit Setup) Click here to exit the installation Click on the computer symbol to continue the installation. Thermal Mass Flowmeter FMT200-ECO2 42/14-41 EN...
In this case, check if you have selected the correct interface, Are all cables connected ? Is the FMT200-ECO2 (Sensyflow eco2) in operation (green light)? Fig. 6-6 If the connection is successfully established, you will see the first page of the configuration program.
6.3.3 Standard specification If the FMT200-ECO2 (Sensyflow eco2) is operating with a standard volume flow measuring unit, you can see the standard state conditions in this window. If the device is running with a mass flow unit, this win- dows stays empty.
Which electrical analog output is active? Analog output: Fig. 6-12 actual measuring value as analog signal (mA or V). Output at error: FMT200-ECO2 signalizes an error at the analog output. Filter factor: time constant of the damping. Offset suppression: Measuring values smaller than the adjusted one will be shown as zero flow.
The control LED still shows permanent red after the procedure. Now, by short interruption of the supply voltage a hardware-reset will be performed. After this, the control LED turns to permanent green and the device works properly. Thermal Mass Flowmeter FMT200-ECO2 42/14-41 EN...
With the standard density of air = 1,293 kg/Nm³ one air-balloon is filled with 0,00905 kg. That means 1 kg/h is approx. 110 air-balloons/h, or 1,84 air-balloons/min. The standard density is related to 0°C, 1013 mbar. With „delete unit“ at the configuration menu, you can delete the new unit. 42/14-41 EN Thermal Mass Flowmeter FMT200-ECO2...
This page allows parameter settings of the analog output. For a better overview, the actual and the new parameters are both shown. The new settings will be sent to the FMT200-ECO2 (Sensyflow eco2) when the save-command is select- ed from the „save data“ menu.
The new settings will be sent to the device when the „save“-command is selected from the „save data“ menu. After saving the data the FMT200-ECO2 (Sensyflow eco2) will initialize itself. During the save-initializa- tion phase, the LED is blinking red.
Changing the configuration 6.5.4 Switch output In this mode the FMT200-ECO2 (Sensyflow eco2) will observe whether the actual measuring value is be- tween the adjusted limits. The polarity informs the eco2, how the signal should react. Fig. 6-20 6.5.5 No output This mode disables the digital output.
Under no circumstances, components such as the honeycomb or sieve may be removed from the probe, bent or damaged in any way. Non-compliance with the cleaning steps described above may lead to damage of the probe. 42/14-41 EN Thermal Mass Flowmeter FMT200-ECO2...
2) It is possible to specify any unit which you can transform into a mass or stan- dard volume flow. For Example 1 ballon/min = 7 l/min - q Thermal Mass Flowmeter FMT200-ECO2 42/14-41 EN...
COMx port of your PC or Laptop to the service and configuration box. Using the configuration software, all measur- ing ranges and output signals can be adjusted to your needs. FMT200-ECO2 RS 232 42/14-41 EN Thermal Mass Flowmeter FMT200-ECO2...
These corrective measures make measurements more expensive and they reduce the ultimate accuracy of the measuring system. FMT200-ECO2 (Sensyflow eco2) measuring systems provide the mass flow rate di- rectly, i.e. without further measurement or correction.
Emission: EN 50 081-1:1992 Immunity: EN 50 082-2:1995 EN 61326-1:1997 Basic standards: EN 61000-4 part 2,3,4,5,6 and 11 ABB Automation Products GmbH Alzenau, Januar 2001 investigationreport: 001-11/98 conformitätsnr.: F01-11-98 42/14-41 EN Thermal Mass Flowmeter FMT200-ECO2...
Page 31
Appendix: Statement of contamination Appendix: Statement of contamination Statement about the contamination of devices and components The repair and/or maintenance of devices and components will only be performed when a completely filled out explanation is present. Otherwise, the shipment can be rejected. This explanation may only be filled out and signed by authorized specialist personnel of the operator.