FEP630, FEH630 ELECTROMAGNETIC FLOWMETER | OI/FEP630/FEH630-EN REV. D
... 9 Diagnosis / error messages
... Error messages
Operation outside of specifications (Out Of Spec.)
Error no. / Range
Text on the LCD display
S065.044 / operation
Inhouse temp. alarm.
Reduce ambient temperature.
S064.041 / operation
EPD alarm.
Secure pipe is completely filled.
S063.040 / operation
TFE alarm.
Secure pipe is completely filled.
S062.039 / operation
Sensor temp. limits alarm.
Change limits or change fluid
temperature.
S061.038 / operation
Conductivity limits alarm.
Change limits or Check
application.
S060.037 / operation
Gas bubble alarm.
Check conditions of application.
S052.016 / operation
Curr.Out 31 / 32 is saturated.
CO process value out of range.
Adapt Qmax.
S051.017 / operation
Curr.Out V1 / V2, V3 / V4
saturated.
CO process value out of range.
Adapt Qmax.
S049.019 / electronics
Option Card 1 com error.
Defective Card. Check Card 1.
Call Service.
S048.020 / electronics
Option Card 2 com error.
Defective Card. Check Card 2.
Call Service.
S047.015 / operation
Pulse output is cut off.
Wrong config. Check pulse out
configuration.
S046.000 / operation
Mass flowrate exceeds limits.
Check flowrate and alarm limits.
S044.001 / operation
Volume flowrate exceeds limits.
Check flowrate and alarm limits.
Cause
Measuring medium or ambient temperature is
outside the spec.
Sensor not filled.
Alarm of the complete filling electrode, but
incorrect, because the sensor is not completely
filled.
The measuring medium temperature is outside
the temperature limit.
The conductivity of the measuring medium is
outside the limit values.
Gas bubbles in the measuring medium
The selected process value of the current output
31/32 is outside the measuring range.
The selected process value of the current output
V1 / V2 or V3 / V4 is outside the measuring range.
Broken hardware of the motherboard or option
card.
EMC interference
Broken hardware of the motherboard or option
card.
EMC interference
The calculated output pulse or the calculated
output frequency is above the configured cutoff
frequency.
The mass flow is below or above the configured
limit values 'Qm Massflow Min' and 'Qm Massflow
Max'.
The volume flow rate is below or above the
configured limit values 'Qv Volumeflow Min' and
'Qv Volumeflow Max'.
Remedy
Check process conditions, reduce temperature.
Check if pipe is empty.
Make sure that the sensor is completely filled.
Check installation and process conditions.
Check process conditions and adjust alarm
threshold.
Check process conditions and adjust alarm
threshold.
Check the process conditions.
Adjust measuring range.
Adjust measuring range.
Check / replace option card in slot 1.
Please contact the service department.
Check / replace option card in slot 2.
Please contact the service department.
Check configuration for the output pulse.
Check the parameterization in menu 'Process
Alarm / ...Alarm Limits' and adjust as needed.
Check volume flow rate.
Check the parameterization in menu 'Process
Alarm / ...Alarm Limits' and adjust as needed.
Check volume flow rate.