FEP630, FEH630 ELECTROMAGNETIC FLOWMETER | OI/FEP630/FEH630-EN REV. D
Digital output 41 / 42, 51 / 52
Can be configured as pulse, frequency or binary output via on-
site software.
A
I
E
41+
012345
51+
42-/
52-
B
I
E
41+
R
B
51+
42-/
52-
A  Digital output 41 / 42, 51 / 52 passive as a pulse or frequency output
B  Passive digital output 51 / 52 as binary output
Figure 48:
(I = internal, E = external, R
Pulse / frequency output (passive)
Terminals
Output 'closed'
Output 'open'
f
max
Pulse width
Binary output (passive)
Terminals
Output 'closed'
Output 'open'
Switching function
Note
Terminals 42 / 52 have the same potential. Digital outputs
DO 41 / 42 and DO 51 / 52 are not electrically isolated from
each other. If an additional electrically isolated digital output
is required, a corresponding plug-in module must be used.
If you are using a mechanical counter, we recommend setting
a pulse width of ≥ 30 ms and a maximum frequency of
f
≤ 30 Hz.
max
R
B
R
B
012345
16...30 V DC
= load)
B
41 / 42, 51 / 52
0 V ≤ U
For f < 2.5 kHz: 2 mA < I
For f > 2.5 kHz: 10 mA < I
16 V ≤ U
CEH
0 mA ≤ I
0.1 to 2000 ms
41 / 42, 51 / 52
0 V ≤ U
2 mA ≤ I
16 V ≤ U
CEH
0 mA ≤ I
Parameterization possible.
See Menu: Input/Output on page 88.
Current output V1 / V2, V3 / V4 (plug-in module)
Up to two additional plug-in modules can be implemented via
the 'Passive current output (red)' option module.
U
R
CE
I
B
A
CE
I
V1+
V2-
U
R
CE
I
B
A Current output V1 / V2, passive
CE
Figure 49: (I = internal, E = external, R
The plug-in module can be used in slot OC1 and OC2.
G11597-02
600
500
400
300
200
100
0
0
≤ 3 V
CEL
< 30 mA
CEL
Permissible source voltage U
< 30 mA
resistance R
where Imax = 22 mA.
CEL
B
≤ 30 V DC
Figure 50: Source voltage for passive outputs
≤ 0.2 mA
CEH
10.5 kHz
Passive current output
Terminals
Output signal
Load R
B
Source voltage U
≤ 3 V
Measuring error
CEL
≤ 30 mA
CEL
Resolution
≤ 30 V DC
*
The source voltage U
≤ 0.2 mA
CEH
an additional area.
B
E
I
R
B
B Current output V3 / V4, passive
= load)
B
5
10
15
20
Uq [V]
for passive outputs in relation to load
q
= Permissible range

*
q
is dependent of the load R
q
39
E
V3+
R
B
V4-
G11897-02
25
30
35
G10323-02
V1 / V2, V3 / V4
4 to 20 mA
250 Ω ≤ R
≤ 600 Ω
B
13 V ≤ U
≤ 30 V
q
< 0.1 % of measured value