Maintenance

Replacing the Printer Paper

The Printer Not Ready message appears on the screen when the paper supply is completely empty. To
avoid the risk of losing patient data, it is better to change the paper roll when the red, end-of-roll, line
appears along the edge of the printout.
If you run out of paper while you are printing patient data, you will be able to restart the printer after you
have put in new paper. No patient data will be lost.
If you replace one type of paper with another, standard with label, for example, be sure to go back to the
second setup screen (Press Help, then Setup, then Page Down) and change the paper type in the printer
options.
Note: It is not necessary to turn off the power to replace the paper. If you do, you will lose any
patient data that is currently stored in the instrument.
To replace the paper
:
1. Push the printer lever on the side of the instrument (see page 1) to the unlocked position and slide the
printer straight out of its housing.
2. Push the paper feed lever on the side of the printer forward to release pressure on the platen.
Printer lever moved to the unlocked position
Paper feed lever
3. Remove the cover on the paper tray and lift out the empty paper roll.
4. Put in a new paper roll, orienting it so that the paper feeds up from the bottom. Feed the paper through
the slot until it comes out the front of the printer.
5. Relock (push back) the paper feed lever.
Zeiss Humphrey Systems Acuitus User Manual
PN 50648 Rev A June 2001
16