15.2
DISMOUNTING THE SAMPLE GAS PUMP FOR EXAMINATION OR MAINTENANCE
For dismounting the sample gas pump N3/5/9 KPE:
W A R N I N G !
•
Disconnect the gas conditioning system from the mains voltage
•
Loosen the 4 screws on the front panel and slightly open it to the front without removing the tubing
and wiring and pull it upwards out of the housing holder.
•
Loosen the 4 fastening screws of the mounting angle on which the pump is mounted (Figure 8)
•
Remove tubing from the pump head
•
Pull out the pump with mounting angle
•
Inspection and maintenance according to operating instructions N3/5/9 KPE (is part of the
supplied manual or can be downloaded from our homepage: www.mc-techgroup.com)
•
After inspection or maintenance, retighten the mounting bracket and mount the front panel
fixing screws
Figure 8
Dismounting the sample gas pump
24
Dangerous voltage. Before any maintenance work is carried out on
the gas conditioning system, disconnect the unit from the mains!
mounting angle