15.3 MAINTENANCE OF THE INTEGRATED PERISTALTIC PUMP TYPE SR25.2

Before starting any maintenance work, make sure that any work done on the device is in compliance
with all relevant regulations and standards.
W A R N I N G !
W A R N I N G !
Flexible tube, conveying belt, contact pulleys and contact springs are the only parts of the pump subject
to wear. They are simple to change.
N O T E !
www.mc-techgroup.com
Inhalation hazard possible, if using toxic or asphyxiant gases!
Purge peristaltic pump with inert gas or air before opening! If the
pump is used for toxic gas or asphyxiant (oxygen-displacing) gas,
it needs to be purged with inert gas or air before opening. Follow
closely all relevant occupational safety regulations during
operation.
Disconnect power supply before opening the device for access.
Make sure that all external power supplies are disconnected.
Aggressive condensate possible!
Media residues in tubing!
Chemical burns caused by aggressive media possible!
Wear protective gloves and protective glasses!
Wear proper protective clothing!
Peristaltic pump is under pressure! Do not open housing!
A peristaltic pump might be part of a system, which is under
pressure. Check pressure before opening peristaltic pump, and
adjust pressure to atmosheric pressure.
If you send back the peristaltic pump to the M&C service for repair,
please let us know what kind of condensate has been pumped.
Before sending the pump back clean all parts from dangerous or highly
aggressive contaminants.