All serviceable parts are mounted in an easily accessible location in the front of the compact gas
conditioning unit CSS-VC...
Change filter elements of the dust and aerosol filters (FPF-0,1GF/FPF+/universal filter/aerosol filter)
when there is too little flow or after visual inspection (see Section 18.1);
Notice
• Check the condensate pump SR25.2 every six months and replace if necessary (see section 18.3.1)
• Clean cooling fins on the compressor cooler of dust regularly using compressed air (see Section
18.4)
18.1

CHANGE THE FILTER ELEMENTS AND O RINGS

To change the filter elements and O-rings:
Unplug the gas conditioning unit from the mains
Unscrew filter glass
Unscrew filter element with filter holder
Exchange filter element and/or O-ring(s). Ensure correct seating of filter element and O-ring(s).
Screw glass filter cap back
18.2

MAINTENANCE OF THE SAMPLE GAS PUMP(S)

Diaphragm and valve plates are the only pump parts subject to wear. They are easy to change. Wear is
indicated by a drop in pneumatic power.
Warnung
Unplug the gas conditioning unit from the mains;
Loosen 4 screws (captive) on the front panel and, without removing the hoses and wiring, bring
slightly forward and pull up out of the housing support; the front panel can now be suspended
back in the housing holder by rotating it 90° (Figure 17).
www.mc-techgroup.com
In order to protect downstream analysers, in the event of a condensate
breakthrough, the wet filter element must always be changed.
Hazardous voltage. Pull the mains plug out before carrying out work on
the gas conditioning unit!
CSS-VC-1.6 | 1.01.03