Description; Figure 3 Design Css-V2 - M&C CSS Series Instruction Manual

19“ housing or wall mounting gas conditioning unit
Hide thumbs Also See for CSS Series:
Table of Contents

Advertisement

9 DESCRIPTION

Figure 3
Design CSS-V2
 Fine filter FPF-2-0,3GF
 Electronic controller
 Flow meter FM40 with flow alarm sensor FA-20mo
 Peristaltic pump SR25.2 with condensate outlet
 Cold appliance socket
 Connection for collective alarm
 Sample gas outlets
 Condensate outlet directly at the peristaltic pump
 Sample gas inlet directly at the heat exchanger
All components of the gas conditioning unit are mounted inside a compact sheet steel case. Filter and
peristaltic pump are mounted to the front panel to ensure easy maintenance.
The minimum flow is determined by the gas sample pump (see also chapter 8) Premature damage can
be caused to the pump membrane if less than the minimal total amount of flow is extracted as a result
of excess pressure.
The gas cooler is equipped with a heat exchanger made of Duran glass, stainless steel SS 316Ti or
PVDF.
The FPF-2-0.3GF fine filter (0.3 µm filter porosity) installed upstream of the N3/5/9 KPE sample gas
pump ensures the necessary solids separation.
14
CSS-V | 1.01.02
www.mc-techgroup.com