15.3.4 CLEANING THE PUMP HEAD

When changing flexible tube or other parts, inspect all parts for dirt before assembling the pump head
and clean them if necessary.
We recommend to clean the parts with a dry cloth. Solvent should not be used, because it can
damage the plastics and synthetic rubber parts. Use oil-free compressed air to clean the parts if
available.
15.3.5 REPAIR INFORMATION FOR INTEGRATED PERISTALTIC PUMP TYPE SR25.2
N O T E !

15.4 CLEANING THE COOLING FINS OF THE COMPRESSOR COOLER

To prevent a decrease of the cooling capacity the cooling fins have to be cleaned from dust periodical.
For this purpose blow compressed air into the ventilation grid at the right side of the housing.
N O T E !
www.mc-techgroup.com
While mounting, make sure that the center of rotation and the roll carrier
(driver) are aligned.
Make sure that the collar of the shaft bore (see Figure 11) faces towards
the front of the pump head while mounting the roll carrier.
Use genuine spare parts only!
Aggressive condensate possible!
Media residues in tubing!
Chemical burns caused by aggressive media possible!
Wear protective gloves and protective glasses!
Wear proper protective clothing!
When sending the peristaltic pump to M&C customer service for repair,
please indicate the type of medium pumped. Before shipping the pump,
please remove hazardous or aggressive contaminations from all parts of
the pump!
With clean cooling fins a DIN A4 sheet of paper is sucked in and
clings to the right side of the housing.
CSS-V | 1.01.02