Ieee-488 Bus Restrictions; Ieee-488 Bus Setup Procedure - Fluke 5790B Operator's Manual

Ac measurement standard
Hide thumbs Also See for 5790B:
Table of Contents

Advertisement

5790B
Operators Manual

IEEE-488 Bus Restrictions

These restrictions apply to all IEEE-488 systems:

IEEE-488 Bus Setup Procedure

To set up the Product on the IEEE-488 bus, choose the address and connect to
a controller. To set up the bus:
1. With the Product off, attach the IEEE-488 cable to the rear panel IEEE-488
2. Push  to turn on the Product.
3. Touch Setup Menu>System Settings>Remote Port.
4. The IEEE-488 bus address for the Product is shown in the GPIB Setup
5. If the Active Remote Port is not GPIB, touch Active Remote Port and then
6. Touch Exit to exit the Remote Port Menu.
5-2
A single IEEE-488 bus system can have a maximum of 15 devices
connected.
The maximum length of IEEE-488 cable used in one IEEE-488 system is
either 2 m (6.56 ft) times the number of devices in the system, or 20 m
(65.61 ft), whichever is less.
connector. Fluke Calibration recommends shielded cables Y8021 (1m),
Y8022 (2m), or Y8023 (4m) from Fluke.
region. To change the address, touch the editable field and enter a new
address with the numerical keys.
touch GPIB.