Linefeed; Ieee-488 Parameters; Ieee-488 Address - Fluke 6050H User Manual

Calibration bath
Hide thumbs Also See for 6050H:
Table of Contents

Advertisement

Linefeed

The final parameter in the serial interface menu is the linefeed mode. This
parameter enables (on) or disables (off) transmission of a linefeed character (LF,
ASCII 10) after transmission of any carriage-return. The linefeed parameter is
indicated by,
Press "SET" to access the linefeed parameter.
= 
The mode may be changed using "UP" or "DOWN" and pressing "SET".
= 

IEEE-488 Parameters

Baths may optionally be fitted with an IEEE-488 GPIB interface. In this case the
user may set the interface address within the IEEE-488 parameter menu. This
menu does not appear on baths not fitted with the interface. The menu is
indicated by,

Press "SET" to enter the menu.

IEEE-488 Address

The IEEE-488 interface must be configured to use the same address as
the external communicating device. The address is indicated by,

Press "SET" to access the address setting.
= 
Adjust the value with "UP" or "DOWN" and then use "SET" to set the address to
the displayed value.
= 
Serial linefeed parameter

Current linefeed setting
New linefeed setting
IEEE-488 parameters menu
IEEE-488 interface address
Current IEEE-488 interface address
New IEEE-488 interface address