Fluke 5790B Operator's Manual page 152

Ac measurement standard
Hide thumbs Also See for 5790B:
Table of Contents

Advertisement

5790B
Operators Manual
SH_LISTALL?
Lists all shunt serial numbers available.
Parameter
Example
SH_LE?
Returns the loading error correction points for the currently selected shunt.
Parameter
Response (comma separated list)
Example
SH_LEADD
Adds a loading error correction for the currently selected shunt.
Parameter
Example
6-62
None
SH_LISTALL?
456123,2A
4562,20MA
1345,20MA
Shows three shunts, a 2 A shunt and two 20 MA shunts and their respective
serial numbers.
None
1. (Integer) Number of data points
2. (Real) Frequency calibration point
3. (Integer) Error in ppm
.
.
.
SH_LE?
5,0.00e+00,0,1.00e+03,0,1.00e+04,0,3.00e+04,0,1.00e+05,0
1. (Real) Frequency calibration point
2. (Integer) Error in ppm
SH_LEADD 2e3,4
Adds a 2.0 kHz data point for the currently selected shunt and sets the
loading error to 4 ppm.