Fluke 5790B Operator's Manual page 119

Ac measurement standard
Hide thumbs Also See for 5790B:
Table of Contents

Advertisement

CAL_DATE
This sets the verification date to the date returned by the internal clock/calendar;
the calibration mode must be unsecured.
Parameter
None
CAL_DATE?
Returns the date of the most recent verification or calibration.
Parameters
No argument (Returns the date of the most recent verification, as set by
CAL_DATE)
ZERO
MAIN
WBND
SERVICE
CAL_DATE? also emulates 5790A syntax with two parameters:
1. ACTIVE, STORED, or OLD (which set the date comes from)
2. One of the following keywords:
DC
AC
ZC
WDC
WAC
ALL
Response
(Integer) Date in the format defined by the DATEFMT command (or in the
utility menus).
Alphabetical List of 5790B Device-Dependent Commands
(Returns the date of the most recent zero calibration)
(Returns the date of the most recent main input adjustment)
(Returns the date of the most recent Wideband input
adjustment)
(Returns the date of the most recent Service adjustment)
(Returns the specified date of DC adjustment)
(Returns the specified date of AC adjustment)
(Returns the specified date of zero calibration)
(Returns the specified date of Wideband gain adjustment)
(Returns the specified date of Wideband flatness
adjustment)
(Returns the specified date of calibration)
6
Remote Commands
6-29