Fluke 5790B Operator's Manual page 140

Ac measurement standard
Hide thumbs Also See for 5790B:
Table of Contents

Advertisement

5790B
Operators Manual
IPADDR ?
Sequential command. Returns the IP address for LAN communication. When
DHCP is enabled, this address will be the address allocated by the DNS server.
When DHCP is disabled, this address will be the entered value of the static IP
address.
Parameter
Response
Example
Will return 129.196.137.45 if DHCP is enabled and the DNS server has
allocated this address to the Product, or will return 129.196.136.119 if DHCP is
disabled and the static address has been previously set to that address.
ISCE0
Loads two bytes into the Instrument Status 1 to 0 Change Enable register
described under "Checking Product Status."
Parameter
Example
Loads 00000000 00110100 (binary) into the register, enabling bits 5
(MCCHG), 4
(MDCHG), and 2 (RNGCHG).
ISCE0?
Reads the contents of the Instrument Status 1 to 0 Change Enable register
described under "Checking Product Status."
Parameter
Response
Example
Returns 52 if bits 5 (MCCHG), 4 (MDCHG), and 2 (RNGCHG) are enabled (set
to 1) in the Instrument Status 1 to 0 Change Enable register.
ISCE1
Loads two bytes into the Instrument Status 0 to 1 Change Enable register
described under "Checking Product Status."
Parameter
6-50
None
String
IPADDR?
The decimal equivalent of the binary number to load into the register. You
can also send a binary, octal, or hexadecimal number if you precede the
number with #b, #o, or #h, respectively. (A number without a preceding
designator is taken as decimal.)
ISCE0 52
None
(Integer) The contents of the Instrument Status 1 to 0 Change Enable register
in decimal.
ISCE0?
The decimal equivalent of the binary number to load into the register. You
can also send a binary, octal, or hexadecimal number if you precede the