SP_SET?
Returns the serial port settings contained in nonvolatile memory.
Response
1. (Integer) One of these baud rate values: 9600, 19200, 38400, 57600 or
115200
2. (CRD) TERM or COMP (Response type)
3. (CRD) XON, RTS, or NOSTALL (Stall method)
4. (CRD) DBIT7 or DBIT8 (Data bits)
5. (CRD) SBIT1 or SBIT2 (Stop bits)
6. (CRD) PNONE, EVEN, ODD, or IGNORE (Parity)
7. (CRD) CR, LF, or CRLF (End-Of-Line)
Example
SP_SET? 9600,TERM,XON,DBIT8,SBIT1,PNONE,CRLF
SRQSTR
Sets the serial/USB/Ethernet remote mode SRQ (Service Request) response
string (up to 40 characters).
Parameter
The serial/USB/Ethernet remote mode SRQ string.
Example
SRQSTR "\nSRQ\n
"\n" denotes the NEWLINE character (hex 0A).
SRQSTR?
Returns the string programmed for serial/USB/Ethernet Mode SRQ responses.
Response
(String) The serial/USB/Ethernet remote mode SRQ string.
STATS
Turns measurement statistics ON or OFF.
If STATS are already on, STATS ON restarts them.
Parameter
1. (Boolean) ON or 1 for on, OFF or 0 for off
Example
STATS OFF
Turns measurement statistics off
STATS ON
Turns measurement statistics on. If measurement statistics were already
enabled, this command will restart the sampling. If measurement statistics
were previously off, this will start sampling if the sample count (see
STATS_COUNT) had previously been set.
"
Note
Alphabetical List of 5790B Device-Dependent Commands
6
Remote Commands
6-67