Fluke 5790B Operator's Manual page 156

Ac measurement standard
Hide thumbs Also See for 5790B:
Table of Contents

Advertisement

5790B
Operators Manual
SP_SET
Sets the serial port settings and saves them in nonvolatile memory.
Parameters
Defaults:
Example
6-66
One or more of the following, in any order:
1. 9600, 19200, 38400, 57600, or 115200 (Baud rate)
2. TERM or COMP (See note below)
3. XON, RTS, or NOSTALL (Stall method)
4. DBIT7 or DBIT8 (Number of data bits)
5. SBIT1 or SBIT2 (Number of stop bits)
6. PNONE, EVEN, ODD, or IGNORE (Parity)
7. CR, LF, or CRLF (End-Of-Line)
The "TERM" or "COMP" parameter sets the responses in serial
remote control to be appropriate for interactive terminal use or
operation under program control. Specifying "TERM" generates
remote responses in plain English for a human operator using a
terminal. Specifying "COMP" generates remote responses for
computer program use. (This command has the same effect as
setting Remote I/F to "TERMINAL" or "COMPUTER" in the RS-
232-C Port Setup Menu.)
9600, TERM, XON, DBIT8, SBIT1, PNONE, CRLF
SP_SET 9600,COMP,XON,DBIT8,SBIT1,EVEN,CRLF
Note

Advertisement

Table of Contents
loading

Table of Contents