Fluke 5790B Operator's Manual page 130

Ac measurement standard
Hide thumbs Also See for 5790B:
Table of Contents

Advertisement

5790B
Operators Manual
CAL_ZERO
Executes dc zeros calibration. An error is generated if the Product cannot make a
measurement (ie, if the short is not applied to the input as specified in the normal
operation section of this SRS). (Overlapped command.)
Parameter
CLOCK
Sets the clock/calendar. If the second parameter is present, the CALIBRATION
security state must be set to off or an execution fault results.
Parameters
Examples:
Sets the clock/calendar to 1:37 p.m., July 17, 2012. (Assuming DATEFMT =
MDY.)
Sets the clock/calendar to 8:00 a.m., March 10, 2012. (Assuming DATEFMT =
DMY.)
CLOCK?
Returns the setting of the clock.
Parameter
Response
Example
Returns
if the clock/calendar is set to 3 p.m., September 1, 2012.
6-40
None
1. Time in 24-hour format as HHMMSS
2. (Optional) Date in the format controlled by the DATEFMT command
CLOCK 133700, 071712
CLOCK 080000, 100312
None
1. (Integer) Time as HHMMSS.
2. (Integer) Date in the format controlled by the DATEFMT command.
CLOCK?
150000,090112