Fluke 55XXA/COIL 50 Instructions Manual

Fluke 55XXA/COIL 50 Instructions Manual

50, 1, 2, and 10-turn current coil

Advertisement

Quick Links

55XXA/COIL 50, 55XXA/COIL 10
Introduction
Use the 55XXA/COIL 50, 55XXA/COIL 10 (the Product or Coil) to calibrate clamp-type current
meters (clamp meters) that operate by two different principles - as current transformers (ac
only), and by the Hall Effect (both ac and dc). For example, it is impractical to calibrate
1000 A-rated current clamp meters with a 1000 A source. However, you can effectively
multiply the current of the current source calibrator by a factor of 50 to support the
calibration and verification of these clamp meters with the 50 turns of the 55XXA/COIL 50.
Contact Fluke
Fluke Corporation operates worldwide. For local contact information, go to our website:
www.flukecal.com.
To register your product, or to view, print, or download the latest manual or manual
supplement, go to our website.
Fluke Corporation
P.O. Box 9090
Everett WA 98206-9090
U.S.A.
+1-425-446-5500
fluke-info@fluke.com
PN 5227455
April 2022 Rev.1, 12/22
(English)
©2022 Fluke Corporation. All rights reserved.
Specifications are subject to change without notice.
All product names are trademarks of their respective companies.
50, 1, 2, and 10-Turn Current Coil
Fluke Europe B.V.
P.O. Box 1186
5602 BD Eindhoven
The Netherlands
Instructions
1

Advertisement

loading
Need help?

Need help?

Do you have a question about the 55XXA/COIL 50 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Fluke 55XXA/COIL 50

  • Page 1 1000 A-rated current clamp meters with a 1000 A source. However, you can effectively multiply the current of the current source calibrator by a factor of 50 to support the calibration and verification of these clamp meters with the 50 turns of the 55XXA/COIL 50. Contact Fluke Fluke Corporation operates worldwide.
  • Page 2: Safety Information

    55XXA/COIL 50, 55XXA/COIL 10 Instructions Safety Information A Warning identifies hazardous conditions and procedures that are dangerous to the user. XW Warning To prevent possible electrical shock, fire, or personal injury: This product is unfused. Do not exceed current and voltage ratings.
  • Page 3: Specifications

    Example: A Fluke Calibration 5560A Calibrator supplies 13 A at 60 Hz to the 55XXA/COIL 50 generating an effective current of 650 A (50 x 13 A), 60 Hz). The 1-year specification of the Calibrator at 13 A is ±(0.07 % + 8 mA), which is equivalent to 0.13 % of the 13 A calibrator output. The total specification of the calibrator and the coil together as a percentage of the output: Specification of calibrator current = ±0.13 %...
  • Page 4 55XXA/COIL 50, 55XXA/COIL 10 Instructions Use the Coil Clamp meters and other current transducers often operate as current transformers, with differing degrees of magnetic coupling between primary and secondary that vary from meter to meter. The position of the clamp meter, with respect to the cable, affects the magnetic coupling between primary and secondary of the current transformer, which causes variation in the reading of the clamp meter.
  • Page 5 50, 1, 2, and 10-Turn Current Coil Use the Coil Figure 2. 55XXA/COIL 50...
  • Page 6: Limited Warranty And Limitation Of Liability

    Instructions LIMITED WARRANTY AND LIMITATION OF LIABILITY This Fluke product will be free from defects in material and workmanship for one year from the date of purchase. This warranty does not cover fuses, disposable batteries, or damage from accident, neglect, misuse, alteration, contamination, or abnormal conditions of operation or handling.

This manual is also suitable for:

55xxa/coil 10