Fluke 5790B Operator's Manual page 123

Ac measurement standard
Hide thumbs Also See for 5790B:
Table of Contents

Advertisement

CAL_INTV?
Returns the calibration interval.
Parameter
None
Response
(Integer) The number of days in the calibration interval (90, 365, 730).
CAL_MODE?
Returns the setting of the CALIBRATION MODE state. (CAL_SW? returns the
calibration security state (secure or unsecure).
Parameter
None
Response
0 = not in service calibration mode, 1 = cal secure mode for SERVICE
CALIBRATION is off
CAL_NEXT
Initiates the next calibration point. An error occurs if the response from
CAL_NEXT? would have been 0,-1 (a series of calibration steps must be begun
using a CAL_AC or CAL_DC or CAL_I2 command). Note that in ac calibration
the frequency is implicit (the proper value can be obtained with the CAL_NEXT?
command).
This initiates the calibration for this step. Make sure the specified
reference is connected and stable before executing CAL_NEXT.
All the same errors that occur with calibration from the front panel can also occur
here (for example if the reference amplitude or frequency is out of tolerance).
Parameter
(Optional. If omitted, the nominal amplitude is used.) A number signifying
reference amplitude (must be within a certain percentage of the expected
amplitude, obtainable using the CAL_NEXT? command.
Alphabetical List of 5790B Device-Dependent Commands
Note
6
Remote Commands
6-33