General Specifications - Fluke 5790B Operator's Manual

Ac measurement standard
Hide thumbs Also See for 5790B:
Table of Contents

Advertisement

5790B
Operators Manual

General Specifications

Warm-up Time ................................................ 30 minutes or twice the time the 5790B has been turned OFF.
Relative Humidity
Operating .................................................... ≤80 % to 30 °C, ≤70 % to 40 °C, ≤40 % to 50 °C
Storage ....................................................... <95 % non-condensing. A power stabilization period of 4 days may be
Altitude
Operating .................................................... 0 - 2000 meters
Non-Operating ............................................ 0 - 12,200 meters
Temperature
Operating .................................................... 0 °C to 50 °C
Calibration ................................................... 15 °C to 35 °C
Storage ....................................................... -40 °C to 70 °C
Electromagnetic Compatibility (EMC)
International ................................................ IEC 61326-1: Controlled Electromagnetic Environment
Korea (KCC) ............................................... Class A Equipment (Industrial Broadcasting & Communication
USA (FCC) .................................................. 47 CFR 15 subpart B. This product is considered an exempt device per
Surge .............................................................. ANSI C62.41-1980, Category A
Reliability ........................................................ MIL-T-2880D, paragraph 3.13.3
Size
Height ......................................................... 17.8 cm (7 in) standard rackmount + 1.5 cm (0.6 in)
Width ........................................................... 43.2 cm (17 in)
Depth .......................................................... 63 cm (24.8 in)
Maximum Power Requirements
5790B ......................................................... 100 VA
Weight
5790B ......................................................... 24 kg (53 lb)
With Wideband ........................................... 24.5 kg (54 lb)
Line Power ...................................................... 50 Hz/60 Hz; 100 V - 120 V, 220 V - 240 V
Safety .............................................................. IEC 61010-1: Overvoltage Category II, Pollution Degree 2
Remote Interfaces .......................................... RS-232, IEEE-488, USB, Ethernet
Confidence Level ............................................ 99 % unless otherwise specified.
DC Zero Cal .................................................... Perform the dc zero calibration every 30 days. In addition, perform the
1-8
required after extended storage at high temperature and humidity
CISPR 11: Group 1, Class A
Group 1: Equipment has intentionally generated and/or uses
conductively-coupled radio frequency energy that is necessary for
the internal function of the equipment itself.
Class A: Equipment is suitable for use in all establishments other
than domestic and those directly connected to a low-voltage
power supply network that supplies buildings used for domestic
purposes. There may be potential difficulties in ensuring
electromagnetic compatibility in other environments due to
conducted and radiated disturbances.
Emissions that exceed the levels required by CISPR 11 can occur
when the equipment is connected to a test object.
Equipment)
Class A: Equipment meets requirements for industrial
electromagnetic wave equipment and the seller or user should
take notice of it. This equipment is intended for use in business
environments and not to be used in homes.
clause 15.103.
IEC 61010-2-030: Measurement 1000 V
dc zero calibration after powering up the unit the first time after
unpacking following a shipment or if exposed to an environmental
change of greater than 5 °C.