Fluke 5790B Operator's Manual page 125

Ac measurement standard
Hide thumbs Also See for 5790B:
Table of Contents

Advertisement

CAL_RPT?
Returns a listing of a calibration report.
Parameters
ACTIVE (The shifts resulting from the last calibration)
STORED (The shifts resulting from the last stored calibration)
CONSTS (A list of all calibration constants)
Response
(IAD) The report followed by the EOFSTR.
CAL_RCSV?
Sequential command. Returns a report in CSV format for a specified calibration
activity (see Chapter 7 for format details).
Parameter
ACTIVE (The shifts resulting from the last calibration)
STORED (The shifts resulting from the last stored calibration)
CONSTS (A list of all calibration constants)
Response
(String)
<formatted report>
CAL_SECURE
Sequential command. Lock/unlocks the calibration security state by use of a
passcode. The passcode is entered as a quoted string of decimal digits (for
example, "5790"). To secure the Calibration state, no passcode is necessary. If
an incorrect password is entered, the Calibration state will automatically be
resecured if it was unsecured.
Parameter
1. ON/OFF
2. <passcode>
Example
CAL_SECURE OFF, "5790"
Unsecures Calibration.
Example
CAL_SECURE ON
Secures Calibration.
Alphabetical List of 5790B Device-Dependent Commands
6
Remote Commands
6-35