Introduction
To avoid electric shock or injury, do not perform the performance tests or
calibration procedures unless you are qualified to do so.
The information provided in this manual is for the use of qualified personnel
only.
The 33X Calibration Information provides the information necessary to verify the performance and adjust
the calibration of the Fluke 333, 334, 335, 336, and 337 ClampMeters, hereafter known as the Meter(s).
The following information is included in this document:
•
Safety Information and International Electrical Symbols
•
Specifications
•
Replacing the Batteries
•
Cleaning
•
Performance Tests
•
Calibration Adjustment
•
User-Replaceable Parts and Accessories
•
Warranty Statement
See the 33x Instruction Cards for complete operating instructions.
Contact Information
To contact Fluke, call:
1-888-99-FLUKE (1-888-993-5853) in USA
1-800-36-FLUKE (1-800-363-5853) in Canada
+31 402-678-200 in Europe
+81-3-3434-0181 Japan
+65-276-6196 Singapore
+1-425-446-5500 in other countries
For additional information about Fluke, its products, and services, visit Fluke's web site at:
www.fluke.com
To register this product, go to
1618765
PN
June 2001
© 2001 Fluke Corporation. All rights reserved. Printed in U.S.A.
WWarning
www.fluke-warranty.com
Calibration Information
®
33X
ClampMeters
1
Need help?
Do you have a question about the 333 and is the answer not in the manual?
Questions and answers
My fluke 334 stopped reading voltage on live circuit is there any reset , I installed brand new batteries
The provided context does not include specific instructions for resetting the Fluke 334 multimeter. However, if the Fluke 334 stops reading voltage on a live circuit, try the following general steps:
1. Turn the rotary switch to OFF.
2. Remove the test leads.
3. Check and replace the battery if needed.
4. Inspect the fuses and replace if blown.
5. Wait a few moments, then turn the meter back on.
If the issue remains, consult the calibration manual or seek service.
This answer is automatically generated