N5290/1A Option 202 (Interconnect Kit N5292-60013) - Keysight Technologies PNA Series Installation Manual

2-port and 4-port microwave network analyzer system (900 hz - 110 ghz / 900 hz - 120 ghz)
Table of Contents

Advertisement

System Installation
PNA, Controller, and Test Head Module Interconnections

N5290/1A Option 202 (Interconnect Kit N5292-60013)

IMPORTANT!
The N5227B and N5247B interconnect kits contain 2.4 mm to 3.5 mm
cables. It is important to connect the 2.4 mm end to the PNA and to
connect the 3.5 mm end to the N5292A test set controller. Improper
connections will result in damage to your instruments.
It is important to use two wrenches when connecting the semirigid cables
to the 1/2/3/4 port 1.85 mm outputs front ports of the PNA to avoid
stressing the connectors and causing damage to your PNA (i.e., the
2.4 mm cable end is compatible with the 1.85 mm connector of the PNA).
Refer to
Figure 3-9
N5290/1A-202—(Interconnect Kit N5292-60013): Install semirigid cables
(N5292-20014 and N5292-20017) and if not done, connect 50 ohm loads
(1810-0118) to the front panel of the PNA and N5292A
3- 14
Figure
3-8.
1818-0118
Keysight N5290/1A Installation Guide

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pna-x seriesN5290aN5291a

Table of Contents