System Installation; Information In This Chapter; Chapter Three At-A-Glance - Keysight Technologies PNA Series Installation Manual

2-port and 4-port microwave network analyzer system (900 hz - 110 ghz / 900 hz - 120 ghz)
Table of Contents

Advertisement

Keysight N5290/1A PNA/PNA-X Series
2-Port and 4-Port Microwave Network Analyzer System
Installation Guide
3

System Installation

IMPORTANT!
A 85059B 1.0 mm Precision Calibration Kit and a 85059V 1.0 mm Precision
Verification Kit are required to complete the installation of the N5290/1A
Millimeter-wave Systems.
In order to maintain compliance to the certificate of calibration it is
recommended that the frequency extenders be connected to the
corresponding test set ports per the Certificate of Calibration provided.

Information in This Chapter

This chapter describes preparing to receiving your 2-port or 4-port N5290/1A
millimeter system and the system's installation procedure(s).

Chapter Three at-a-Glance

Section Title
"Getting Prepared" on page 2
"Getting Assistance from Keysight"
on page 3
"Receiving the System" on page 4
"PNA, Controller, and Test Head
Module Interconnections" on
page 6
"Initializing the N5290/1A
Broadband System" on page 22
"Running the Installation
Calibration" on page 23
"Troubleshooting the N5293/5A
Millimeter Modules" on page 26
Summary of Content
Discusses how to download the service guide, tools required for
installation, and the time required to install the system.
How to contact Keysight for support with your system.
What to do when you receive your system, verifying the parts received,
and customer engineering support.
A discussion and procedures for assembling your N5290/1A system.
Discusses how to initialize the N5290/1A system.
Describes the N5290/1A millimeter wave installation calibration.
Describes the types of error messages that might be seen, if a
millimeter wave module is incorrectly installed.
3-1