Table 1-6
Power Requirements of the System
Standard Equipment
Instrument
N5222/27/42/47B
N5292A millimeter head controller
N5293A millimeter head (x2)
N5295A millimeter head (x2)
Total (N5222/27/42/47B)
Values are based on 120 Vac supplied to each instrument at 60 Hz.
The N5292A millimeter head controller supplies power to the test head
modules.
1-18
Maximum VA Rating
350
210
(powered from controller)
(powered from controller)
560
Keysight N5290/1A Service Guide
Maximum BTU/hour
1195
717
(powered from controller)
(powered from controller)