Toroidal Conductivity - ABB AWT210 Operating Instruction

2-wire conductivity, ph/orp pion transmitter
Hide thumbs Also See for AWT210:
Table of Contents

Advertisement

34
AW T 2 1 0 | 2 - W I R E C O N D U C T I V I T Y, P H/O R P, P I O N T R A N S M I T T E R | O I/A W T 2 1 0 - E N R E V. C
...11 Configuration
... Calibrate

Toroidal conductivity

Menu
Calibrate
Exit
Select
Menu
Comment
PV Zero Cal
PV Span Cal
Temperature Cal
Hold Output
HART versions only. Not available if HART in Multidrop or Fixed Output mode.
Disabled/Enabled.
When Enabled: output is held automatically at the beginning of a PV calibration and released automatically at the end.
When Disabled: output remains active throughout any PV calibration.
Edit Cal
PV Slope
PV Offset
Temp Slope
Temp Offset
Restore Defaults
Used to calibrate the sensor.
Note. Calibrate menus are sensor-specific – the menus in this section are applicable to
transmitters connected to toroidal conductivity sensor types.
Access to the Calibrate menu is permitted via the Calibrate and Advanced levels or directly from
an Operator page using the Cal button.