Device Setup

Menu
Device Setup
Exit
Select
Menu
Comment
Security Setup
HW Write Prot .
Read-only.
Disabled/Enabled
SW Write Prot .
Disabled/Enabled
Calibrate Password Available only at Advanced access level.
Advanced Password Available only at Advanced access level.
HART Login
Enabled for HART versions only.
Disabled/Enabled
PDM Compatibility
Available for HART versions only.
Enable this option if the device is connected to Siemens Simatic PDM.
Disabled/Enabled
Reset To Defaults
Available only at Advanced access level.

Display

Menu
Display
Exit
Select
Menu
Comment
Operator Page 1
Selects range of values displayed on Operator Page 1 – see page 51 for examples of character formats/sizes.
PV/PV + Temperature/PV + Temp + PV%/PV + Temp + O/P
Operator Page 2
Selects range of values displayed on Operator Page 2 – see page 51 for examples of character formats/sizes.
PV/PV + Temperature/PV + Temp + PV%/PV + Temp + O/P
Contrast
Adjust contrast of display.
Language
Selects the language to be displayed on screen:
English/Deutsch/Français/Español/Italiano
Note. Language options are displayed with localized spelling.
Used to setup security parameters, set PDM compatibility (for HART-enable transmitters) and to
reset default (factory-set) parameters.
Used to setup Operator pages (1, 2), select the display language and set the display contrast.