AW T 2 1 0 | 2 - W I R E C O N D U C T I V I T Y, P H/O R P, P I O N T R A N S M I T T E R | O I/A W T 2 1 0 - E N R E V. C

6 Electrical installation

DANGER
• If the transmitter is used in a manner not specified by the
Company, the protection provided by the equipment may
be impaired.
• Refer to page 7 for electrical installation
considerations in Hazardous areas.
• The transmitter conforms to Installation Category II of IEC
61010.
• All equipment connected to the transmitter's terminals
must comply with local safety standards (IEC 60950,
EN61010-1).

Terminal connections

Communication module connections
Figure 15 Connections overview
HART modules
Power
Ammeter
+
+
1
2
3
4
5
HART modules
A B
A B
1
2
3
5
4
DANGER – CONNECTION/CABLE
REQUIREMENTS
• The connection terminals accept cables with peripheral
wire cross-section of:
– min.: 0.14 mm
(26 AWG)
2
– max.: 1.5 mm
(14 AWG)
2
• Do not use a rigid conductor material as this can result in
wire breaks.
• Ensure the connecting cable is flexible.
• To ensure the sensor cable length is sufficient, allow an
additional 100 mm (4 in) of cable to pass through cable
glands into the housing.
• Ensure the correct connections are made to suit the
transmitter variant.
Sensor module connections
pH/ORP/plon modules
6
7
1
2
3
4
5
2-electrode conductivity modules
1
2
3
4
5
6
7
4-electrode conductivity modules
1
2
3
4
5
6
7
Toroidal conductivity modules
1
2
3
4
5
6
7
8
8
8
8
15

Advertisement

Table of Contents
loading

Table of Contents