Sensor Setup - ABB AWT210 Operating Instruction

2-wire conductivity, ph/orp pion transmitter
Hide thumbs Also See for AWT210:
Table of Contents

Advertisement

AW T 2 1 0 | 2 - W I R E C O N D U C T I V I T Y, P H/O R P, P I O N T R A N S M I T T E R | O I/A W T 2 1 0 - E N R E V. C

Sensor Setup

pH
Menu
Sensor Setup
Exit
Select
Menu
Comment
Measurement Type
Selects a measurement type compatible with the connected sensor:
pH/ORP/Ion Conc/pION
pH Sensor Type
Enabled only if Measurement Type = pH.
Select a measurement type compatible with the connected sensor:
Glass/Antimony/Custom
PV Unit
Enabled only if Measurement Type = Ion Conc.
Sets the process value units.
%/ppm/ppb/µg/l/mg/l
Isopotential Pt .
Isopotential point.
Enabled only if Measurement Type = pH and pH Sensor Type = Custom.
Asymmetric Pot .
Asymmetric potential.
Enabled only if Measurement Type = pH and pH Sensor Type = Custom.
Valence
The ion valence determines the millivolt change per decade of concentration.
Enabled only if Measurement Type = Ion Conc.
–3/–2/–1/1/2/3
Magnitude
The number of magnitudes (ranging from 1 to 3) that define the ion
concentration output.
Enabled only if Measurement Type = Ion Conc.
1/2/3
End Magnitude
The ion concentration state functions by associating an end millivolt value to an end magnitude value.
Enabled only if Measurement Type = Ion Conc.
10/100/1000
End mV
The end millivolt value associated with the end magnitude value.
Enabled only if Measurement Type = Ion Conc.
Temp . Unit
Temperature unit displayed and applied to all temperature parameters/values.
Not enabled if Temperature Compensation Type = Manual.
°F/°C
Temp . Comp . Type
Temperature compensation type – see page 51.
Manual/Auto/Auto Solution
Note. When Auto is selected and a temperature sensor is recognized, the temperature compensation calculation
is based on the Nernst equation detailed on page 51. When Auto Solution is selected and a temperature
sensor is provided with the probe, the calculation is based on the Nernst solution coefficient equation detailed
in detailed on page 51. When Manual is enabled/selected, the value set in the Manual Temp. parameter is used
for the referenced temperature compensation.
Manual Temp .
Temperature value used for compensation when Temp Comp. Type = Manual compensation mode.
Enabled only if Temperature Compensation Type = Manual.
pH Soln . Coeff .
Enabled only if Measurement Type = pH and Temp. Comp. Type = Auto Solution.
mV Soln . Coeff .
Enabled only if Measurement Type = pH and Temp. Comp. Type = Auto.
Temp . Sensor Type
Read-only.
3K Balco 2-wire/3K Balco 3-wire/PT 100 2-wire/PT 100 3-wire/PT 1000 2-wire/PT 1000 3-wire/None
Detect TC Type
TC Detection Status
Read Only/Not Recognized/Recognized
Used to access standard setup parameters.
Note. Sensor Setup menus are sensor-specific – the menus in this section are applicable to
transmitters connected to pH sensor types.
35