ABB AWT210 Operating Instruction page 48

2-wire conductivity, ph/orp pion transmitter
Hide thumbs Also See for AWT210:
Table of Contents

Advertisement

48
AW T 2 1 0 | 2 - W I R E C O N D U C T I V I T Y, P H/O R P, P I O N T R A N S M I T T E R | O I/A W T 2 1 0 - E N R E V. C
...13 Specification
Configuration security
Hardware switch configuration
Access to configuration is enabled only after a hardware
switch has been set. This switch is situated behind a
tamper-evident seal.
Password protection
Access to configuration levels is enabled only after the
user has entered a password:
• Calibrate level: user-assigned password
• Advanced level: user-assigned password
• Service level: factory set password
Input
pH/ORP/plon sensor types
pH:
Glass, Antimony (Sb)
ORP: (Redox): Platinum (Pt), Gold (Au)
pIon: Custom user-programmable
Input impedance
>1x10
13
pH/ORP/plon measurement range and resolution
Type
Range
0 to 14 pH
pH
(–2 to 16 over range)
ORP
–1500 to 1500 mV
pIon
–1500 to 1500 mV
Dynamic response
<1 second for 90 % step change at 0 seconds damping
Damping
Configurable: 0 to 99.9 seconds
Conductivity sensor types
AWT210: ABB 2-electrode conductivity sensors
AWT210: ABB 4-electrode conductivity sensors
AWT210: ABB toroidal conductivity sensors
Conductivity measurement range and resolution
AWT210 2-electrode conductivity transmitter:
Cell
Conductivity
constant
0.01
0 to 200 µS/cm
0.1
0 to 2000 µS/cm
1
0 to 20000 µS/cm
Display
resolution
repeatability
0.01 pH
1 mV
1 mV
Display
range
resolution
repeatability
0.001 µS/cm
measurement
0.01 µS/cm
0.1 µS/cm
...Conductivity measurement range and resolution
AWT210 4-electrode conductivity transmitter:
Sensor
group
A
B
AWT210 toroidal conductivity transmitter:
Sensor
ABB
toroidal
Temperature input
Temperature element types
Pt100 (2 or 3-wire)
Pt1000 (2 or 3-wire)
3k Balco (2 or 3-wire)
None
Measurement range and resolution
Accuracy
Temperature
element
Pt100
±0.01 pH
Pt1000
±1 mV
3K Balco
±1 mV
None
pH/ORP/plon temperature compensation modes
Type
Manual
pH
ORP
pIon
Conductivity temperature compensation modes
Temperature
element
0 to 15 % NaOH
Accuracy
0 to 20 % NaCI
0 to 18 % HCl
±1.0 % of
0 to 20 % H2SO4
range per
Pure water neutral salt
decade
Pure water trace base
Pure water trace acid
User-defined
Conductivity
Display
range
resolution
0 to 2000 mS/cm
0.1 µS/cm
0 to 2000 µS/cm
0.01 µS/cm
Conductivity
Display
range
resolution
0 to 2000 mS/cm
1.0 µS/cm
Automatic temperature compensation
Automatic temperature compensation
Automatic temperature compensation
Manual temperature compensation
Temperature
range
–20 to 200 °C
(–4 to 392 °F)
User-programmable
20 to 300 °C (–4 to 572°F)
Nernstian
Automatic
with solution
Nernstian
coefficient
AWT210
AWT210
2-electrode
4-electrode
Accuracy
repeatability
±0.5 % of
measurement
range per decade
Accuracy
repeatability
±0.5 % of
measurement
range per decade
Accuracy
Repeatability
±0.1 °C (±0.18 °F)
after calibration
N/A
Solution
compensation
coefficient
AWT210
toroidal