Panel Mounting (Optional) - ABB AWT210 Operating Instruction

2-wire conductivity, ph/orp pion transmitter
Hide thumbs Also See for AWT210:
Table of Contents

Advertisement

AW T 2 1 0 | 2 - W I R E C O N D U C T I V I T Y, P H/O R P, P I O N T R A N S M I T T E R | O I/A W T 2 1 0 - E N R E V. C

Panel mounting (optional)

Referring to Figure 13:
1 Cut the correct sized hole in panel
2 Insert the transmitter into the panel cut-out
3 Screw one panel clamp anchor screw
bracket
D
until 10 to 15 mm (0.39 to 0.59 in) of the thread
protrudes from the other side of the bracket and position
one clamp
E
over the end of the thread.
NOTICE
The correct torque is critical to ensure proper compression
of the panel seal and achieve the IPX6/NEMA 4X hosedown
rating – see step 6.
Dimensions in mm (in)
30 (1.2)
Panel cut-out dimensions
B
Figure 13 Panel mounting the transmitter
.
A
.
B
C
into the left-hand
138
+1.0
–0.0
+0.04
(5.43
)
–0.0
A
30
(1.2)
A
4 Holding assembly
left-hand recess on the rear of the transmitter and secure
with bracket securing screw
washer remains in the position fitted.
5 Repeat steps 3 and 4 for the right-hand panel clamp
assembly.
6 Torque each panel clamp anchor screw to 0.5 to 0.6 Nm
(4.42 to 5.31 lbf/in).
+1.0
138
–0.0
+0.04
(5.43
)
–0.0
E
D
F
together, position bracket
F
. Ensure that the plastic
G
25.5
(1.00)
73 (2.87)
Mounting panel –
max. thickness
G
6 (0.24)
C
13
into the
D
Gasket

Advertisement

Table of Contents
loading