Input/Output; Diagnostics - ABB AWT210 Operating Instruction

2-wire conductivity, ph/orp pion transmitter
Hide thumbs Also See for AWT210:
Table of Contents

Advertisement

40
AW T 2 1 0 | 2 - W I R E C O N D U C T I V I T Y, P H/O R P, P I O N T R A N S M I T T E R | O I/A W T 2 1 0 - E N R E V. C
...11 Configuration

Input/Output

Menu
Input/Output
Exit
Select
Menu
Comment
Eng . Low
pH devices – cannot be edited if Measurement Type = Ion Concentration.
Conductivity devices cannot be edited if Measurement Type = Concentration.
See Easy Setup (page 28) or Sensor Setup (page 35) levels for the associated sensor type.
Eng . High
pH devices – cannot be edited if Measurement Type = Ion Concentration.
Conductivity devices cannot be edited if Measurement Type = Concentration.
See Easy Setup (page 28) or Sensor Setup (page 35) levels for the associated sensor type.
Damping
Fault Current
HART only.
High/Low
Output Type
HART only.
Linear/Function-Generator
Function-gen Table
HART only.
Enabled only if Output Type = Function-Generator.
X1/Y1/X2/Y2/X3/Y3/X4/Y4/X5/Y5/Table Error
None
Not Saved
Trim 4mA
HART only.
Not available if HART in multidrop or fixed output mode.
Trim 20mA
HART only.
Not available if HART in multidrop or fixed output mode.
Loop Test
HART only.
Not available if HART in multidrop or fixed output mode.

Diagnostics

Menu
Diagnostics
Exit
Select
Menu
Comment
Sensor Diagnostics
pH, 2-electrode conductivity, 4-electrode conductivity only.
Disabled/Enabled
Reference Imp . Limit
pH only. Enabled only if Sensor Diagnostics = Enabled.
Disabled/Enabled
Diagnosis Status
Read-only.
Actual/Simulated
Used to configure input/output values.
See page 21 for all diagnostic messages/actions.