Pv Span Calibration - ABB AWT210 Operating Instruction

2-wire conductivity, ph/orp pion transmitter
Hide thumbs Also See for AWT210:
Table of Contents

Advertisement

46
AW T 2 1 0 | 2 - W I R E C O N D U C T I V I T Y, P H/O R P, P I O N T R A N S M I T T E R | O I/A W T 2 1 0 - E N R E V. C
...12 Calibration
...Toroidal conductivity sensor calibration

PV Span calibration

Apply span and wait for
1
stable reading
Ensure that a span solution is
present at the sensor, monitor the
process value and continue (
the next step once the value has
stabilized.
2
Enter the new value
Enter the desired PV value by
pressing the
key to move the
cursor and the
keys to
change the value. When the new
value has been entered press the
key to continue.
3
Sampling
The procedure moves automatically
to the next stage once the PV has
been sampled.
4
Completion
Following a successful calibration
the calculated coefficients are
displayed.
Following an unsuccessful
calibration the reason for failure
is displayed.
– – –
PV Zero Cal
PV
752 uS/cm
Continue When Stable
) to
Abort
Continue
– – –
PV Zero Cal
PV
752 uS/cm
0
752
New
Continue
Next
– – – –
PV Zero Cal
PV
752 uS/cm
Abort
– – –
PV Zero Cal
PV
752 uS/cm
PV Slope
99.69%
PV Offset
0.00 uS...
Exit
– – –
PV Zero Cal
PV
752 uS/cm
Calibration Failed
Offset Too High
Exit