ABB AWT210 Operating Instruction page 50

2-wire conductivity, ph/orp pion transmitter
Hide thumbs Also See for AWT210:
Table of Contents

Advertisement

50
AW T 2 1 0 | 2 - W I R E C O N D U C T I V I T Y, P H/O R P, P I O N T R A N S M I T T E R | O I/A W T 2 1 0 - E N R E V. C
...13 Specification
Approvals, certification and safety
Factory Mutual (FM) Intrinsic Safety
Available with polycarbonate & aluminium enclosures
Intrinsic Safety
• Class I, Div 1, Group A, B, C, D, T4
• Class II, Div 1, Group E, F, G, T4
• Exia
Enclosure type/ingress protection classification
• 4X*/IPX6
Ambient temperature range
• –25 °C =< Ta =< 60 °C
Factory Mutual (FM) Non-incendive
Available with aluminium enclosure only
Non-incendive
• Class I, Div 2, Group A, B, C, D, T4
• Class II, Div 2, Group F, G, T4
• Class III
Enclosure type/ingress protection classification
• 4X*/IPX6
Ambient temperature range
• –25 °C =< Ta =< 60 °C
Canadian Standards Authority (CSA) Intrinsic Safety
Available with polycarbonate & aluminium enclosures
Intrinsic Safety
• Class I, Div 1, Group A, B, C, D, T4
• Class II, Div 1, Group E, F, G, T4
• Exia
Enclosure type/ingress protection classification
• 4X*/IPX6
Ambient temperature range
• –25 °C =< Ta =< 60 °C
Canadian Standards Authority (CSA) Non-incendive
Available with aluminium enclosure only
Non-incendive
• Class I, Div 2, Group A, B, C, D, T4
• Class II, Div 2, Group F, G, T4
• Class III
Enclosure type/ingress protection classification
• 4X*/IPX6
Ambient temperature range
• –25 °C =< Ta =< 60 °C
ATEX Intrinsic Safety
Available with polycarbonate & aluminium enclosures
Intrinsic Safety
• II 1G Ex ia IIC T4 Ga when used with appropriate barriers
Ingress protection classification
• IPX6
Ambient temperature range
• –20 °C =< Ta =< 60 °C
IECEx Intrinsic Safety
Available with polycarbonate & aluminium enclosures
Intrinsic Safety
• II 2G Ex ia IIC T4 Ga when used with appropriate barriers
Ingress protection classification
• IPX6
Ambient temperature range
• –20 °C =< Ta =< 60 °C
EMC
Emissions and immunity
Meets requirements of IEC61326 for an industrial
environment.
*
4X Hosedown self-assessed not approved by 3
rd
party.
DS/AWT210-EN Rev. A