8 Maintenance
8.1 Care, cleaning, disinfection and sterilization
52
8.1.5.1

CEREC Bluecam without sapphire glass

8.1.5.1.1
General information
The CEREC Bluecam is a very sensitive optical device and must
therefore be handled with the utmost care. Protect the front lens and the
prism against scratching and clean them with a clean lint-free cloth and
ethanol (commercially available cleaning alcohol) whenever a film is
noticed during imaging.
Dental health care providers are advised to select the reprocessing
method that aligns with their infection control process. An effective
infection control program is practical, reasonable, and reproducible.
Two methods have been validated to reprocess the CEREC Bluecam
without sapphire glass between patient care:
● High-level disinfection of the prismatic tube without sapphire glass 
[→ 55]
● Dry heat sterilization of the prismatic tube without sapphire glass 
[→ 58]
One of these methods is strongly recommended to be used for the
reprocessing of the prismatic tube.
8.1.5.1.2
Components of the CEREC Bluecam without sapphire glass
A
Press detent to release
B
Prismatic tube
C
Prism without sapphire
glass (bCL / bCB)

Sirona Dental Systems GmbH

Operating Instructions CEREC AC
D
Camera support
6 pcs., Order No. 5945360
E
Protective cap
F
Front lens