Invacare Storm 4 Series Service Manual page 69

Hide thumbs Also See for Storm 4 Series:
Table of Contents

Advertisement

1. Remove rear and centre shroud. See 6.8.1 Replacing
Rear Shroud, page 55 and 6.8.2 Replacing Centre
Shroud, page 55).
2. Pull batteries out of battery box. See 6.10.3 Making
Batteries Accessible, page 75.
3. Check fuse cable (1) and battery cable (2) for visible
damage and crushing points.
4. Replace any damaged cables.
5. Check all other cables for visible damage and crushing
points. Replace any damaged cables.
6. Pull each plug coming out of power module carefully.
The plug should not come out of socket.
7. If one of plugs is loose, press it back into socket with
light pressure. Plug must engage.
8. Make sure that plug is now firmly in socket. If not,
repeat previous operation.
9. Refit parts in reverse order.
10. Test all functions.
Installing Lifter Column Cables
If a type B lifter column is installed, see Installing on Lifter Column Type B, page 70.
Fig. 6-61 Lifter Column Type A
4 mm Allen key
Oblique pliers
Tie wraps
1533060-N
Fig. 6-59
Fig. 6-60
Fig. 6-62 Lifter column type B
Service
69

Advertisement

Table of Contents
loading

This manual is also suitable for:

Storm 4 x-ploreStorm 4

Table of Contents