Invacare® Storm⁴ Series
The programming device shows the following
GB MOTOR CALIBRATION Chair will drive! Are
wheels raised? EXIT YES
GB MOTOR CALIBRATION -BEGIN- to start.
Wheels will drive! EXIT BEGIN
GB MOTOR CALIBRATION in progress Please
wait.
GB MOTOR CALIBRATION Successful! EXIT
3. Separate programming device from mobility device. Calibration is complete.
Replacing Relay
Phillips screwdriver size 2
3 mm Allen key
4 mm Allen key
8 mm socket wrench
Before replacing the relay, make it accessible.
1. If possible, use lifter to move seat to uppermost
position.
2. Turn off controls on remote.
3. Remove seat plate or sling seat (1). See Modulite
manual Adjusting seat width (Modulite seat).
4. Remove shroud (2).
36
You should enter this
"YES" (if drive wheels have been raised)
"BEGIN" (if drive wheels have been raised)
No entry required. Wait till end of calibration.
"EXIT"
Fig. 6-21
1533060-N