GE XMTC User Manual page 32

Thermal conductivity binary gas transmitter
Table of Contents

Advertisement

Chapter 3. Operation and Programming
3.6.1 Perform Cal (cont.)
Click on Zero Field Cal to calibrate the zero value, or on Span Field Cal to calibrate the span value. In either case, a
window similar to Figure 13 below opens.
Figure 13: Field Cal Execution Window
Click Yes to calibrate, or Abort Field Cal to stop the calibration and return to the previous menu. The result of a
completed calibration is shown in Figure 14 below.
Figure 14: Field Cal Results Window
Click on Previous Item or on Next Item/Enter to return to the previous window, or on Exit Page to return to the
Instrument Menu.
22
XMTC User's Manual