Chapter 5. Specifications

5.1 Performance

Accuracy
Linearity
Repeatability
Stability
Response Time
Measurement Ranges
(typical)
Measurement Gases
(typical)
Ambient Temperature Effect
Required Sample Gas
Flow Rate
Required Reference Gas
Flow Rate
XMTC User's Manual
±2% of span
±1% of span
±0.5% of span
Zero: ±0.5% of span per week
Span: ±0.5% of span per week
20 seconds for 90% of step change
0-2%
0-5%
0-10%
0-25%
0-50%
0-100%
50-100%
80-100%
90-100%
95-100%
98-100%
H
in N
air or CO
2
2,
2
He in N
or Air
2
CO
in N
or Air (minimum range 0-20% CO
2
2
SO
in Air (minimum range 0-10% SO
2
Ar in N
or air (minimum range 0-20% argon)
2
±0.05% of span per °C
0.1 to 4.0 SCFH (5 to 2000 cc/min), 0.5 SCFH (250 cc/min) nominal
0.01 to 4.0 SCFH (5 to 2,000 cc/min), 0.5 SCFH (250 cc/min) nominal
)
2
)
2