Simulation; Device Reset - Siemens sitrans LR200 Instruction Manual

2-wire, 6 ghz pulse radar level transmitter
Hide thumbs Also See for sitrans LR200:
Table of Contents

Advertisement

Simulation

Note: The Simulation parameter influences output to the control system.
Simulate Analog Input to AIFB1 or AIFB2
Allows you to input a simulated value in order to test the functioning of the Analog Input
Function Blocks.
1.
Open the menu Device – Simulation, and select the desired function block.
Simulation (Measured Value)
2.
Click on Simulation (Measured Value).
3.
Enable simulation, enter a value, and click
on Transfer. The LCD displays the
substitute value.
4.
After simulation is complete, disable
simulation.
Simulation (Output)
2.
Click on Simulation (Output).
3.
Select Target Mode MAN, enter a value, and click on Transfer. The LCD displays the
substitute value.
4.
After simulation is complete, reset Target Mode to AUTO to disable simulation.
Simulate Input
Allows you to simulate the sensor value which is input to the Level Transducer Block. This
tests everything between the Level Transducer Block and Output.
1. Open the menu Device – Simulation, and select Simulation (Input).
2. To enable simulation select Fixed or Ramp.
If you select Ramp, enter the step length and number of steps.
3. Enter the simulated value and click Transfer.
4. After simulation is complete, disable simulation.

Device Reset

Reset options
Factory Defaults
Standard Defaults
Informational
Functional
Warm Start
Reset Address to 126
7ML19985JR01
SITRANS LR200 (PROFIBUS PA) – INSTRUCTION MANUAL
Result
Resets all parameters excluding device addresses to the manufacturer's
default settings.
Resets all parameters excluding device addresses to the PROFIBUS default
settings
Resets parameters such as Tag.
Resets parameters that control device behavior, such as Low Calibration Pt.
Has the same effect as recycling power to the device
• Resets the PROFIBUS device address to 126
• If the address lock was on, will disable the lock.
AIFB 1
[
PERCENT
40.61
SUBSTITUTE VAL.
]
Page 41

Advertisement

Table of Contents
loading

Table of Contents