4.18.6. Total Calibration Interval
Note: The device always operates in years. Changing the units affects only
the parameter view of the Calibration Interval parameters in SIMATIC PDM.
User-configurable recommended time between product calibrations.
Values
a)
Units are selectable only via SIMATIC PDM.
4.18.7. Maintenance Status
Displays the status of the Maintenance limits.
In SIMATIC PDM, open the menu View – Device Status, click on the
Maintenance tab and check the Calibration Schedule Status window.
4.18.8. Acknowledge Status
Acknowledges the status of the Maintenance limits.
In SIMATIC PDM, open the menu View – Device Status, click on the
Maintenance tab and check the Calibration Schedule Status window.
4.18.9. Acknowledge
Acknowledges the current limit status.
To acknowledge an alert via SIMATIC PDM:
a) Open the menu View – Device Status and click on the Maintenance tab.
b) In the Calibration Schedule Status section click on Acknowledge
Warnings.
To acknowledge an alert via the handheld programmer:
a) Press RIGHT arrow
Mode.
b) Press RIGHT arrow
Page 88
SITRANS LR200 (PROFIBUS PA) – INSTRUCTION MANUAL
a)
Units
: hours, days, years
Range: 0 to 20 years
Default: 1.0 year
twice to open parameter view and activate Edit
to acknowledge the alert.
7ML19985JR01