Firmware Upgrade Instructions, continued
Procedure
(continued)
D-6
3.13. Select the Measurement Module radio button and click on Continue to
proceed. The following screen will appear.
3.14. Select the file to upload. Click on Browse and select filename
MeasCode.b4u from the files supplied by Dranetz-BMI.
3.15. When done selecting the file to upload, click on Continue and wait until
the following screen appears.
3.16. Click on Continue to proceed.
3.17. If a Calibration Failed page is displayed, repeat the process from Step
3.11.
NOTE: If several attempts fail, turn the power switch to OFF then turn
back to ON, and repeat the process from Step 3.11.
Continued on next page