4.4.3 Printhead pressure control
Evaluate the pressure necessary for extruding the biomaterial. Extrusion pressure depends mainly
on material viscosity.
1.
Set printhead pressure. (Pressure should be between 0-200 kPa with an internal air
compressor and 0-700 kPa with an external air compressor.)
2.
Test bioink flow - Press the droplet button to assess the bioink flow.
There is one droplet button for each printhead.